Meter Maintenance and Troubleshooting
Not enough blood or control
solution was drawn into the test
strip for measurement or was
applied after the test had
started.
Discard the test strip and repeat
the blood glucose or control
test.
An electronic error occurred, or in rare cases, a
used test strip was removed and reinserted.
Remove the batteries, press any button, and
reinsert the batteries. Perform a blood glucose or
control test.
84
The blood or control drop was
applied to the test strip too late.
Discard the test strip and repeat
the blood glucose or control
test.
The temperature is above or below the proper
range for the system.
Refer to the test strip package insert for system
operating conditions. Move to an area with the
appropriate conditions, wait 5 minutes, and
repeat the blood glucose or control test. Do not
artificially heat or cool the meter.
Blood or control solution was
applied to the test strip before
Apply drop appeared.
Discard the test strip and repeat
the blood glucose or control
test.