LaMotte 2020t/i Mode D'emploi page 30

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

11. Fill the tube to the fi ll line with the
turbidity standard. Pour the standard
down the inside of the tube to avoid
creating bubbles. Cap the tube.
12. Wipe the tube thoroughly with a lint-
free cloth.
13. Open the meter lid. Insert the tube
into the chamber. Align the index line
on the tube with the index arrow on
the meter. Close the lid.
ENTER
14. Press
to select Scan Sample
and scan the standard. The screen
will display Reading for about 1
second. The result will appear on the
screen.
"Overrange" will be displayed if
the reading is out of range. Dilute
the sample or select a mode that
is appropriate for the range of the
sample.
15. Press
to scroll to Calibrate.
Turbidity WB (N)
0.99
NTU
Scan Blank
Scan Sample
12:00:00
001/500
Turbidity WB
0.99
NTU
Scan Sample
Calibrate
12:00:00
001/500
30

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

1974-t1974-i

Table des Matières