LaMotte 2020t/i Mode D'emploi page 29

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.
Rinse a clean tube (0260) three
times with the blank. If samples are
expected to read below 1 NTU the
meter should be blanked with a 0
NTU Primary Standard or prepared
turbidity-free (<0.1 NTU) water. For
the most accurate results, use the
same tube for the blank and the
sample.
6.
Fill the tube to the fi ll line with the
blank. Pour the blank down the inside
of the tube to avoid creating bubbles.
Cap the tube.
7.
Wipe the tube thoroughly with a lint-
free cloth.
8.
Open the meter lid. Insert the tube
into the chamber. Align the index line
on the tube with the index arrow on
the meter. Close the lid.
ENTER
9.
Press
to select Scan Blank
and scan the blank. The screen
will display Blank Done for about
1 second and then return to the
Turbidity - With Blank Menu.
10. Rinse a clean tube (0260), or the
same tube, three times with the
turbidity standard.
Turbidity WB (N)
Scan Blank
Scan Sample
12:00:00
001/500
29