SAMPL'AIR LITE
Dimensions
:
Weight :
Noise :
Sampling head :
Airflow :
Sampling mode:
Battery :
Battery charger :
NOTE: qualified personnel only should carry out replacing batteries
10.1 Correspondence table : sampling time/volume sampled
Average speed
10.2 Environmental conditions for good operational practice :
The running of Sampl'air Lite is ensured under the following conditions:
•
When operating : Temperature within 15°C and 35°C
When charging battery :Temperature within 15°C and 35°C
•
•
Hygrometry within 30% and 85% of relative humidity non condensing
Atmospheric pressure within 700 and 1040 HPa
•
•
According to the type of lighting used (strong or indirect natural light, light bulbs, neon tubes), the
action of the infrared remote control may be reduced.
•
Connectors are protected from corrosive gas or liquid with folding caps.
!
!
!
After unpacking or after storage at low temperature, it is strongly
recommended waiting at least 2 hours before using the device (to let it
acclimatize to room temperature)
The company is not responsible for any damage due to failure to operate the
device without following the
56
MANUEL UTIILSATION / USER'S MANUAL
10. TECHNICAL FEATURES
130 x 170 x 300 mm
1,8 kg
50 dBA at 1 meter
Stainless steel AISI 316L, autoclavable, 258 holes diameter 0,7 mm
100 litres / mn, ± 5%
time (1 to 99 minutes) or volume (50 to 9999 litres)
NiMH - 9,6V - 3,8 Ah, autonomy : 25 cycles of 10 min
(Ambient temperature higher or equal to 20°C, backlight deactivated)
FRIWO MPP30 FW7362/15
Necessarily ordered from BIOMÉRIEUX S.A
Input : 100 – 240 V~ / 50 – 60 Hz / 700 mA
Output : 15 V= / 2 A ; Low Voltage (SELV)
Interchangeable mains plugs :
EN 60320 type, France / Europe double pole, outlet plugging
-
-
IEC double pole type, use with a cord
TIME
30s
1 mn
2 mn
4 mn
8 mn
10 mn
: - 16,8 m/s for 100 l/min
instructions.
VOLUME
50 l
100 l
200 l
400 l
800 l
1m³
MA107501