6.2 Setting the delay

When using the Sampl'air Lite for the first time, no is set (the delay is 00:00)
If you want a delay time before starting sampling:
SAMPL'AIR
<
DELAY
Enter
DELAY
00:00
Key in + Enter
DELAY 10:00
SAVE ?
Enter
Delay time is modified
The countdown timer allows you to enter a time value that will be displayed on the start confirmation screen.
Count down starts as soon as you have pressed the Enter key. When the time is up, the collection starts.
A sampling cycle is started from the Sample menu.
SAMPLING FLOW RATE
!
!
!
The Sampl'air Lite flow rate is set for 100 litres / minute, +/- 5 minutes (in normal
operating conditions).
16.8m/s.
The grid (258 holes diam. 0.7mm) has been developped in association with the best
specialists of aerodynamic modelling and underwent cycles of technical and
operational validation in real conditions.

7.1 Sampling programs

When using the Sampl'air Lite for the first time, two programs are displayed
These programs may be deleted and replaced by programs created by the operator (up to 10 sampling
programs.)
To select the sampling program you want to start, use the directional arrows and confirm your choice by
pressing Enter.
MANUEL UTIILSATION / USER'S MANUAL
1- Select the 'Delay' menu using the directional arrows, then
press Enter
>
2- Modify
Modify the seconds, confirm by pressing Enter (range from
00:00 to 59:59)
The delay time is cancelled by dialling 00:00
3- Press Enter to confirm delay time ?,

7 ACHIEVING A QUALITY SAMPLING

With 100l/min, the impaction velocity onto the agar is
1000 L
10 MIN
the
minutes,
and
DELAY
Back to the
screen
MA107501
confirm
by
pressing
Enter.
43