SAMPL'AIR LITE
8.5 Troubleshooting
Problem
No display
Battery charger is connected,
the red warning light does
not light up
The charger is connected, the
green charging led lights up,
the red charging led does not
light up.
Screen display :
EMPTY BATTERY
Screen display :
LOW BATTERY
Screen display :
unsuitable Power supply
Le Sampl'air Lite is switched
on, but some keys do not
work.
Screen display :
ERREUR 11
Screen display :
ERROR 62
Screen display :
ERROR 63
MANUEL UTIILSATION / USER'S MANUAL
Possible cause
st
1
case : the battery is empty or
the charger is not connected
nd
2
case: The battery is charged and
the charger is unplugged. Electronic
problem
st
1
case : the charger of the battery
is not correctly connected
nd
2
case : the voltage supplied by
the charger is not correct
rd
3
case : electronic problem
st
1
case : the battery is already
charged
nd
2
case : an error message is
displayed
rd
3
case : the battery is not
charged, any error message is
displayed
Battery is empty
Charge with low battery
The voltage supplied by the battery
charger is <13V or >17V.
rst
1
case : Program is jammed
2nd case : keyboard is out of
order or electronic problem
Electronic problem : battery
circuit breaking
Turbine's speed variates
abnormally
Electronic problem : Tachymètre
turbine
Corrective action
Charge the Sampl'air Lite (§ 8.3)
Contact the after-sales service
Check the connection of the charger
Use a battery charger adapted: 15Vcc, 2A
minimum
Contact the after-sales service
the Sampl'air Lite can be used
See below for the meaning of the error
message
Contact the after-sales service
put the Sampl'air Lite on charge (§ 8.3)
At the end of the sampling cycle, put the
Sampl'air Lite on charge (§ 8.3)
Disconnect immediately the battery
charger, use a battery charger delivering
15Vcc, 2A mini. (§ 8.3)
Press the "forced stop" button (§3.2)
Contact the After Sales Service
Contact the After Sales Service
Make sure that the air circulates
properly, that the dish is placed
correctly.
If everything seems OK, contact the
After Sales Service (electronic
problem).
Contact the After Sales Service
MA107501
53