8. Operation
Turn the unit on by means of the main
switch (33) at the back of the unit. By
this, the scanner (1) at the bottom and
the display (2) at the front of the unit
will be activated. An acoustic signal
shows that the unit is ready for opera-
tion; the display will show the name of
the machine Drufomat scan.
8.1 Scan modus
Every outer package of Dreve thermo-
forming blanks is labelled with a bar-
code label.
This barcode contains the encrypted
product name, material strength as
well as the individual heating and cool-
ing times.
In order to scan the material data with
the Drufomat scan, the barcode is to
be placed centrally and evenly under
the scan field (1).
36
There will be an acoustic signal for
the confirmation of the data capture
and the display will show all neces-
sary data.
Info
The light emanating from the
scanner is harmless. Even in case
of direct eye contact it has no neg-
ative impact!
After 15 min. without use the
scanner turns off automatically.
By pressing any key it can be re-
activated again.
During an active thermoforming
process the scanner is deactivated.
After termination of the process
it will be re-activated automati-
cally.

8.2 Pressure adjustment

The pressure of the Drufomat scan is
The arrow to the left of the bar chart
controlled digitally and can individually
shows the currently adjusted pressure.
be adapted to the thermoforming blank
The arrow can be moved along the bar
in question between 29 and 87 psi
upwards and downwards with the +/-
infinitely variable.
keys (e/i). The pressure can also be mod-
It is possible to control and to modify the
ified during an ongoing thermoforming
adjusted pressure and the actual pressure
process!
during the thermoforming process via
the bar chart (g) of the display.
Info
The minimum operating pressure
is 29 psi (2 bars). If a thermoform-
ing process is interrupted before
having reached the 29 psi (2 bar)
mark (indication of bar chart (g)
in the display), the piston will be
ventilated and will return to its
original position.

8.3 Start of programme

In order to start the selected pro-
Swivel the plate reception inwards at the
gramme, put a blank that corresponds
handle (10) just up to the mechanical
to the entered barcode resp. program-
interlock in the centre of the unit. Also
ming onto the plate reception (11) and
swivel in the heating element (4) at the
fix it by means of the tension ring (13).
corresponding handle (5) just up to the
The model can either be put onto the
mechanical interlock in the centre of the
nub plate of the F insert (6) or fixed by
unit. Close the ventilation valves of the
taking out the nub plate and embedding
tension ring (12) and the F insert (6).
it into granules directly in the F insert.
Press the OK key (f). There will be an
acoustic signal as confirmation for the
Info
activation of the heating; the scanner
For an optimum adaptation of the
will turn off and the adjusted heating
blank towards the model always
time will elapse in the display. As safety
place the models in a way that the
and identifying feature for an active
anterior teeth are directed towards
heating there will be four LED alight at
the centre of the blank.
the front bottom edge of the housing.
When using hard blanks it is recom-
mended to embed the models into
granules. Models that are placed
onto the nub plate should not
exceed a max. height of 23 mm.
37