The Percent Configurations are defined below (defaults in Bold).
Configure Item
Option(Bold is default)
Auto Tare
Off/On
Accumulate
Off/Manual/Automatic

4.4.3 Input/Output (I/O) Setup

The I/O's setup can be customized for various user preferences.
The I/O's setup is defined below (defaults in Bold).
Item
Enable
Note: The I/O's will only work when the I/O Option board has been installed. See the accessory list in
section 9.4 for information. The option I/O board provides two isolated inputs and four dry-contact
normally open relay outputs which can be used for simple process weighing.
Press the button
With the Softkey corresponding to the icon
down the list and highlight Application Mode.
Enter this sub-menu by pressing the button
corresponding to the icon
In the Application Mode menu, enter the Percent
sub-menu.
Available Settings
On, Off
to enter the Main Menu.
.
Description
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Off: The Softkey "∑" doesn't display.
Manual: The Softkey "∑" displays. User can press
the key to accumulate the stable weight.
Automatic: The Softkey "∑" displays. The weight
will be accumulated automatically.
Note:
The load to be accumulated has to be >= 10d and
the next accumulation can only start once the pan
has been cleared.
When LFT is ON (no such limit when LFT is OFF
or the approved model is OIML),
a. gross weight and net weight cannot be
accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated
at the same time;
b. a new sample can be accumulated after
reaching gross 0 or net 0.