Defender 5000 Indicators
The Check Configurations are defined below (defaults in Bold).
Configure Item
Option (Bold is default)
Check Mode
Weighting/Counting
Auto Tare
Off/On/Accept
Accumulate
Off/Manual/Automatic
Audible Signal
Off/Under and
Over/Accept/Under/Over
Off/On
Auto Opt.
Opt. Beep
Off/On
APW Auto Save
Off/On

4.3.4 Input/Output (I/O) Setup

The I/O's setup can be customized for various user preferences.
The I/O's setup is defined below (defaults in Bold).
Item
Enable
Discrete Input 1
Discrete Input 2
Discrete Output 1
Discrete Output 2
Discrete Output 3
Discrete Output 4
Note: The I/O's will only work when the I/O Option board has been installed. The option I/O board
provides two isolated inputs and four dry-contact normally open relay outputs which can be used for
simple process weighing.
Available Settings
On, Off
Off, Zero, Tare, Clear Tare, Print, Unit, Accumulate
Off, Zero, Tare, Clear Tare, Print, Unit, Accumulate
Off, Under, Over, Accept, Under/Over
Off, Under, Over, Accept, Under/Over
Off, Under, Over, Accept, Under/Over
Off, Under, Over, Accept, Under/Over
Description
Weighing: Check weighing mode.
Counting: Check counting mode.
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Accept: If the object weight is in the range of the Over and
Under Limit you set, auto tare will be performed.
Off: The Softkey "∑" does not display.
Manual: The Softkey "∑" displays. User can press the key
to accumulate the stable weight.
Automatic: The Softkey "∑" displays. The weight will be
accumulated automatically.
Note:
The load to be accumulated has to be >= 10d and the next
accumulation can only start once the pan has been cleared.
no such limit when LFT is OFF or the
When LFT is ON(
approve mod is OIML)
,
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross
0 or net 0.
Off: No beep.
Under and Over/Accept/Under/Over: Beep when
reaching the selected check point.
Off: Auto Opt. is off.
On: The APW will be optimized automatically when doing
count weighing.
Off: Opt. Beep is off.
On: When the APW is optimized, the beeper will beep
once.
Off: APW Auto Save is off.
On: If the APW is get from sampling, and a counting
library is selected, the new APW will save to library after
optimized.