Span Calibration - OHAUS Defender 5000 TD52P Mode D'emploi

Indicateurs
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Defender 5000 Indicators
Long press the button
Menu. Press the Softkey corresponding to the
icon
to enter the Calibration sub-menu.
Zero Calibration is on the top of the list of
Calibration by default. Just press the Softkey
corresponding to the icon
Calibration.
Clear the pan and then press the Softkey
corresponding to the icon
The message 'Completed' will be displayed on
the screen.
Exit Zero Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon

3.3.2 Span Calibration

Span Calibration uses one point. The span calibration point is established with a calibration mass
placed on the scale.
Note: Span Calibration should be performed after Zero Calibration.
Calibration procedures:
Long press the button
Menu.
Press the Softkey corresponding to the icon
enter the Calibration sub-menu.
Scroll to Span Calibration using the Softkey
corresponding to the icon
to enter the Main
to initiate Zero
.
.
.
to enter the Main
to
.
EN-19