Check Weighing (Default); Check Counting - OHAUS Defender 5000 TD52P Mode D'emploi

Indicateurs
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Defender 5000 Indicators
Two different modes can be selected in the Check Mode: Weighing, Counting.
Setup check limits according to section 4.3.1 or 4.3.2. Place object on the pan to check if the weight
is within the limits.

4.3.1 Check Weighing (default)

Make sure that the Check mode is set to Check
Weighing in the configuration menu.
Place objects on the pan. The Under/Accept/Over
status is shown in the progress bar area while the
actual weight of the item is shown on the main
Display Line.
Defining Over/Under Limits
Press the button Edit Limit to define the limit for
weighing.
Select Over or Under Limit and press the button
corresponding to the icon
The Check Configurations are defined below (defaults in Bold).
Configure Item
Option(Bold is default)
Check Mode
Check Weighting/ Check
Counting
Auto Tare
Off/On/Accept
Accumulate
Off/Manual/Automatic
Audible Signal
Off/Under and
Over/Accept/Under/Over

4.3.2 Check Counting

to edit the value.
Description
Check weighing mode
Check counting mode
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a
container automatically.
Accept: If the object weight is in the range of the Over and
Under Limit you set, auto tare will be performed.
Off: The icon "∑" does not display.
Manual: The icon "∑" displays. User can press the key to
accumulate the stable weight.
Automatic: The icon "∑" displays. The weight will be
accumulated automatically.
Note:
The load to be accumulated has to be >= 10d and the
next accumulation can only start once the pan has been
cleared.
When LFT is ON(
the approve mod is OIML)
a. gross weight and net weight cannot be accumulated;
b. gross 0 has to be reached before a new sample
accumulation.
When LFT is OFF,
a. gross weight and net weight can be accumulated;
b. a new sample can be accumulated after reaching gross
0 or net 0.
Off: No beep.
Under and Over/Accept/Under/Over: Beep when reaching
the selected check point.
no such limit when LFT is OFF or
,
EN-41