8. Repairs
WM 8296
WM 5274
WM 8812
WM 12093
WM 93061
WM 21086
You can perform the following operations on
the STATION MEDUMAT with bracket and
on the Wall bracket STATION MEDUMAT:
Replacing carrying handle, WM 5274:
1. Undo the two securing screws.
Note that there are two washers per
screw.
2. Remove the old carrying handle and
position the new one in the appropriate
drill holes.
3. Secure the new carrying handle with the
screws supplied.
Replacing plastic cap, WM 8812:
1. Pull the loose plastic cap off the spring
plate.
2. Apply a little adhesive to the inside of the
new plastic cap. We recommend
Loctite 601.
English