Table 4 Color measurement options
Options
More
Save icon
Send Data icon
Color Scale
View Graph
View Table
View Values
Scale & Units
Ph.Eur.: Auto
Color Scale Selection
Illuminant/Observer: C/2°
Color Scales for Operator ID
Description
For further Options
Symbol: Store Data, if Instrument Setup > Data Log Setup > Auto Store: Off is selected.
Symbol: Recall Data, if Instrument Setup > Data Log Setup > Auto Store: On is selected.
To send data to a printer, computer or USB memory stick (USB A) or network.
Select the color scale
VIEW GRAPH shows the spectral graph of the transmittance or absorbance graph.
Note: View Graph is activated after the first measured value.
VIEW TABLE displays the spectral transmittance values T% from 380 nm to 720 nm.
VIEW VALUESdisplays the result of the last color calculation.
UNITS: Select absorbance or transmittance.
SCALE: In the automatic Scale mode, the y-axis is automatically adjusted so that the total scan is displayed.
The manual Scale mode allows sections of the scan to be displayed.
European Pharmacopoeia
Select AUTO or the REQUIRED SCALE
Define 3 different color scale combinations with up to 3 different color scales for the result display.
Selection 1: Klett Color No. + Yellowness Index
Selection 2: Klett Color No. + ADMI Color No. + Yellowness Index
Selection 3: Yellowness Index + individual transmittance
Illuminant: Select C, A or D65
Observer: 2° or 10°
Individual selection of the color scales for the operator ID.
English 98