Atlantic Therapeutics Innovo Mode D'emploi page 27

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Rated Outputs – Voltage/Currents for Stress urinary incontinence Programme 1
Parameters
Output RMSA
Output RMSV
Output Frequency
DC Component
Pulse Width t1
Max Output Current (peak)
Rated Outputs – Voltage/Currents for Urge urinary incontinence Programme 2
Parameters
Output RMSA
Output RMSV
Output Frequency
DC Component
Pulse Width t1
Max Output Current (peak)
Time required to reach operating temperature is 1 hour.
Electrical Specifications:
Operating Range:
Transport & Storage Range:
If required, always seek assistance to set up, use or maintain your INNOVO
should report any unexpected operation or event to Atlantic Therapeutics.
Accessories
Under no circumstances should anything other than Atlantic Therapeutics accessories be used with your
®
INNOVO
controller (Type 207). Any others may not be compatible with your controller and could degrade the
minimum safety levels. You can purchase all accessories from Atlantic Therapeutics:
®
INNOVO
Shorts Size XS
®
INNOVO
Shorts Size S
®
INNOVO
Shorts Size M
®
INNOVO
Shorts Size L
®
INNOVO
Shorts Size XL
Neck Strap
Lead
Battery Pack
Conductive Spray (50ML)
Conductive Spray (250ML)
Important:
Under no circumstances should anything other than the correct type of battery pack (rechargeable batteries 1 x
7.2V
(6 x 1.2 AAA) volt (NiMH) DC battery pack) be used with your controller. These can be purchased from the Atlantic
Therapeutics careline.
Power Supply (2504-0806)
Input: 100-240V ~ /50-60Hz/400mA
Output: 12V DC / 1.0A
Ktec AC Adaptor, model KSAS0121200100D5
Important: Do not use any other chargers or power supplies. The power supply is included as part of this
medical electrical equipment. To isolate from the mains, the charger/power supply plug should be removed from
the socket.
Expected Service Life: 2 years.
500
1K
30.3mA
24.4mA
15.1V
24.4V
50HZ
50HZ
0V
0V
620µS
620µS
120mA
100mA
500
1K
5.2mA
5.1mA
2.6V
5.1V
10HZ
10HZ
0V
0V
250µS
250µS
70mA
70mA
Temperature:
Humidity:
Atmospheric Pressure:
Temperature:
Humidity:
Atmospheric Pressure:
1005-0807
1005-0808
1005-0809
1005-0810
1005-0810
1220-0826
1200-0807
3924-0312
1500-0802
1500-0809
1.5K
16.6mA
25.0V
50HZ
0V
620µS
70mA
1.5K
4.5mA
6.75V
10HZ
0V
250µS
70mA
5 to 40°C
15 to 93% RH (non-condensing)
70 to 106kPa=
-25 to +70° C
10 to 93% RH (non-condensing)
50-106kPa
®