Hach QuikChem 8500 2 Série Mode D'emploi page 3

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Table of contents
Specifications
on page 3
General information
on page 4
Installation
on page 7
Additional information
Additional information is available on the manufacturer's website.
Specifications
Specifications are subject to change without notice.
Specification
Dimensions bench space (W x D x H) 2.72 x 0.83 x 0.58 m (9 x 2.75 x 2 ft) (recommended)
Location
Pollution degree/installation category
Protection class
Operating temperature
Storage temperature
Altitude limit
Certifications
Warranty
System unit (5 channels)
Dimensions (W x D x H)
Weight
Power requirements
Fuse rating
Lamp
Sample processing module (SPM)
Dimensions (W x D x H)
Weight
Power requirements
Fuse rating
Pump (12 position)
Dimensions (W x D x H)
Weight
Details
Laboratory (indoor use only)
II
I
5 to 40 °C (41 to 104 °F), maximum 80% relative humidity non-
condensing for temperatures up to 31 ºC and decreases linearly to 50%
relative humidity at 40 ºC
–20 to 70 °C (–4 to 158 °F)
Up to 3000 m (10,000 feet) above sea level
CE: IEC/EN 61010-1 per LV Directive (2006/95/EC), IEC/EN 61326 per
EMC: Directive (2004/108/EC)
cETL: CSA C22.2 No 61010-1
ETLus: UL 61010-1
1 year
70.5 x 71.2 x 26.7 cm (27.7 x 28.0 x 10.5 in.)
46.5 kg (102.51 lb)
100–115 V ±10%, 230 V ±10%; 50–60 Hz, 800 W
T, 8 A, 100–115 V, T, 4 A, 230 V
Tungsten/Halogen, 12 V, 50 W, MR16
69 x 9 x 26 cm (27.2 x 3.5 x 10.2 in.)
4 kg (8.8 lb)
115 VAC, 1.2 A, 60 Hz or 230 VAC, 0.8 A, 50 Hz
T, 1.6 A, 250 V, 5 mm x 20 mm
15 x 26 x 26.7 cm (5.9 x 10.2 x 10.5 in.)
5.6 kg (12.3 lb)
English 3