After having disinfected the accessories, shake them vigorously and place them on a
paper towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
STERILISATION
Accessories C1.1-C1.2-C1.3-C1.4-C2-C3-C4-C5 can be sterilised.
Device: Fractionated vacuum overpressure steam steriliser in accordance with EN 13060.
Implementation: Wrap every single part to be treated with a sterile barrier system or
packaging in accordance with Norm EN 11607. Place the packed components in the steam
steriliser. Run the sterilisation cycle according to the operating instructions of the device by
selecting a temperature of 134°C and a time of 10 minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile barrier
system or packaging.
The sterilisation procedure is validated in its conformity to ISO 17665-1.
At the end of each use store the device complete with accessories in a dry place away
from dust.
AIR FILTERING
The device is quipped with an extraction filter (A3) that must
be replaced when it is dirty or changes colour. Do not wash or
reuse the same filter. Regular filter replacement is necessary to
help ensure proper compressor performance. The filter must be
checked regularly.
To replace the filter:
Insert a slotted screwdriver between the edge of the filter and the body. Lift the filter and
extract it by rotating and pulling up. The filter was designed to always remain secured in its
seat. Do not replace the filter during use.
Only use original accessories and spare parts by Flaem, we disclaim any liability in
the event of using non original spare parts or accessories.
AVAILABLE SPARE PARTS
Description
- RF6 Plus nebuliser kit composed of:
nebuliser, mouthpiece with valve and nasal prong
- 1 m connection tube
- Mask Set composed of: adult mask, pediatric mask and elastic band
- Spare air filter kit (2 pcs)
15
Code
ACO368P
ACO35
ACO462P
ACO164P