Cleaning and Disinfection

WARNING
To prevent cross-contamination, always
follow the disinfection instructions in this
IFU.
WARNING
To avoid material damage and injury, clean
and disinfect according to this IFU:
EN
• No other chemicals are allowed.
• Never clean with chlorine.
• Chlorine will deteriorate the surface of
the material.
WARNING
To avoid injury, always remove the sling
before disinfecting the lift.
WARNING
is only intended for "single patient use".
All Sara Plus
cleaned when it is soiled or stained and between
Close all buckles and hook and loop attachment on
Check the symbols on the product label for
If allowed according to the product label, tumble
Do NOT
wash with other items that have rough surfaces
or sharp objects
use any mechanical pressure, pressing or rolling
during the washing and drying process
use bleach
use gas sterilization
use autoclave
dry clean
steam
ironing
Flites, should be
12
Cleaning Instruction for the
with Chlorine based solution, up to 10 000 ppm, or
Flites
Flites
Flites has been subjected to any such
Flites
(See Fig. 25 )
If the Flites
(See Fig. 26 )

Cleaning Chemicals

Flites) use common
softener, iodine based disinfectants, bromine and

Disinfection

or disinfect the Flites