• Do not move the cuff or put it on if the control unit is
switched on.
• Only connect the connection cable for the electrode plug
to the intended control unit and the cuff that is included.
• Do not pull on the connection cables during treatment.
• Do not bend or pull on the end of the connection cable.
• Do not wear any electronic devices, such as watches,
while using the device.
Precautions
Warning
• During the initial few minutes, use the device while sit-
ting or lying down to minimise the risk of accidental
injuries as a consequence of isolated cases of vagal
responses (feeling of faintness). If you feel faint, switch
off the device immediately, lie down and support the
legs in an elevated position (approx. 5 – 10 min).
• The treatment should be comfortable. If the device does
not work properly, or you feel unwell or experience pain,
stop using it immediately.
• Only remove the cuff once the device has been switched
off.
• Do not use the device near (~ 1 m) shortwave or micro-
wave devices (such as mobile phones) as these may
cause fluctuations in the output values of the device.
• The use of the device may be limited in the presence
of electromagnetic disturbances. This could result in is-
sues such as error messages or the failure of the display/
device.
• Avoid using this device directly next to other devices or
stacked on top of other devices, as this could lead to
faulty operation. If, however, it is necessary to use the
device in the manner stated, this device as well as the
other devices must be monitored to ensure they are
working properly.
• The use of accessories other than those specified or
provided by the manufacturer of this device can lead to
an increase in electromagnetic emissions or a decrease
in the device's electromagnetic immunity; this can result
in faulty operation.
• Failure to comply with the above can impair the perfor-
mance of the device.
• Never immerse the device in water or other liquids.
• Do not use in the vicinity of highly flammable substances,
gases or explosives.
Notes on the electrodes
Important
• The electrodes must not be attached to areas of broken
skin.
• Max. recommended output value for electrodes is
5 mA/cm
.
2
• Effective current densities over 2 mA/cm
increased attentiveness from the user.
21