Acyclic Services; Cyclic Services - Siemens SIPAN 32 Instructions De Service

Masquer les pouces Voir aussi pour SIPAN 32:

Publicité

Les langues disponibles

Les langues disponibles

4.8.10

Acyclic Services

4.8.11

Cyclic Services

Dispositif de mesure de conductivité SIPAN 32
C79000-B5477-C045-02
MPIP sarl - 94, rue Mohamed Tounsi - 16013 - ROUIBA - ALGERIE
Tél./Fax : 021 81 11 22 - GSM : 070 91 75 15 ou 072 900 654 - E-mail : mpip@mpi-dz.com
The diagnostic object consists of 4 bytes.
Byte
Bit
Meaning when "1"
0
0
Electronic temperature
too high
1
--
2
--
3
--
4
Memory error
5
Measurement failure
6
7
Init Error
Byte Bit
Meaning when "1"
1
0
--
1
--
2
Configuration invalid
3
Restart
4
Coldstart
5
Maintenance
6
--
7
Ident_Number violation
The basic English texts originate from the device data file (GSD,
see section 4.8.16 "GSD (device data file")
Bytes 2 and 3 are reserved for future extensions.
Using the acyclic services you can read every readable parameter of
the PROFIBUS profile or write every writable parameter with the
appropriate access authorization.
The parameters of the PROFIBUS and their attributes (read and/
or write; see section 4.8.17 "Object List").
The control system requires the device--specific device data file for
connection setup.
SIPAN 32 with PROFIBUS PA operates with the standard device data
file according to the device profile for analyzers. The device data file
can be loaded via Internet or a CD (see section 4.8.7.1.2.
"Operation/control with control system").
All profile files and those of the Siemens devices are already stored in
the SIEMENS control system.
Mise en service et fonctionnement
Electronic temperature
high
Memory error
Measurement failure
Initialization error
Configuration invalid
Restart
=
Cold start
Maintenance request
Violation
of ident_number (GSD)
4-57