C 6010 and C 6012 decomposition vessels are experiment autoclaves
and must be tested by a professionally trained person each time before
they are used.
An individual application is understood here to mean a series of experiments that
are performed under roughly the same conditions in terms of pressure and tem-
perature. Experiment autoclaves must be operated in special chambers (C 6000
gs/iso).
Repeated tests:
The decomposition vessel must be subject to repeated tests (internal tests and
pressure tests) by a person with professional training. The intervals between
tests must be determined by the operator based on experience, operating manner
and the material used in the decomposition vessel.
The declaration of conformity loses its validity if mechanical modifica-
tions are made to the experiment autoclaves or if stability can no longer
be guaranteed as a result of heavy corrosion (e.g. pitting by halogens, see
section "Halogens C 6012").
The threading on the body of the decomposition vessel and cap screw in
particular are subject to a high level of mechanical stress and must there-
fore be monitored regularly for wear and tear.
The condition of the seals must be checked for functionality must be ensured by
means of a test for leaks (see sections "Leakage test").
Pressure tests and service tasks on the decomposition vessel must only be per-
formed by persons with professional training.
We recommend that the decomposition vessel be sent into our factory
for inspection and repairs if necessary after either 1000 experiments or
after one year or, depending on the application, even sooner than this.
Definition of person with professional training:
A person with professional training as defined in these operating instructions is
someone:
1. whose training, knowledge and experience gained through practical activities
ensures that that person will perform the tests in a proper manner.
2. who is sufficiently reliable.
3. who is not subject to any instructions in terms of testing activity.
4. who is equipped with suitable testing equipment if necessary.
5. who can provide suitable proof demonstrating compliance with the require-
ments listed in 1.
Operating pressure containers:
National regulations and laws for operating pressure containers must be observed!
Anyone who operates a pressure container must maintain it in proper condition,
must monitor it and perform necessary maintenance and repair tasks without
delay, and must take measures appropriate for the circumstances to ensure safety.
A pressure container must not be operated if it exhibits defects that
could endanger those working with it or third parties.
You can obtain a copy of the pressure vessel regulation from Beuth Press.
Use only original spare parts from IKA

Included with delivery C 6010/C 6012

1 x Decomposition vessel
1 x Quartz dish C 4
1 x Operating instructions
1 x Emergency-vent knob
1 x SW 8/10 double open ended wrench
2 x SW 5,5/7 double open ended wrench
1 x SW 17 Wrench
5 x Kanthal ignition wire (C 6010)
2 x Platinum ignition wire (C 6012)
1 x O-ring set
1 x C 710.4 cotton threat
1 x Test report
!
®