Drive Unit; Assembly Of Clamp And Stand; Working With The Unit - IKA T 10 basic Mode D'emploi

Disperser
Masquer les pouces Voir aussi pour T 10 basic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Drive unit

The open up a wide range of possibilities for dispersion technolo-
gy in the conventional laboratory with a respective performance
output of approximately 75 watts at 30.000 1/min.
The speed of the T 10 basic dispersion unit set by using the
differently adjusting wheel.

Assembly of clamp and stand

Overview (fig. 1 and fig. 2)
Assembly
• Clamp (5) as in fig. 1 represented over the dispersion equip-
ment push, until it engages in the position planned for it
audibly.
• Tighten the turning handle (6) by hand up to the notice.
• Screw the clamp (5) (fig. 1) by means of cross sleeve onto the
stand.
For safe working the drive units are fastened by means of a cross
sleeve (7) to the plate stand (8). Pay attention when the assemb-
ling of the dispersion equipment at the stand to a perfect and firm
seat of the cross sleeve.
In order to increase the stability of the mechanical structure, the
drive unit must be installed as closely as possible to the stand sea-
weeds.
Disassembly
• For the disassembly of the dispersion equipment of the
clamp (5), first the turning handle (6) is to be loosened.
• Hold the dispersion equipment with a hand, with which other
hand now the dispersion equipment knows through pressures
on the locking lever (fig. 2) from the clamp is solved.
14

Working with the unit

Overview (fig. 1)
Check that the voltage listed on the designation plate matches the
mains voltage. Also pay attention to the ambient conditions listed
in the technical data.The power socket used must be earthed
(protective earth conductor contact).
Assembly of the dispersion tool
- Open the shaft lock (3) by turning to the left as far as the stop.
- Insert the dispersing element into the collet as far as the stop
(you will feel it click into place after some resistance).
- Lock the shaft lock (3) by turning it to the right until it noticeably
clicks into place. The marks on the shaft lock (3) and on the casing
should be over each other when closed.
The distance between the dispersion tool and the vessel bottom
should not be less than 10 mm.
Please observe also the operating instructions of the dispersion
tool to ensure a safe usage.
The entire unit may also be arranged somewhat off centre in order
to avoid any undesired air admission caused by the strong rotatio-
nal turbulence.
Set the adjusting wheel (2) to the minimum speed before swit-
ching on the appliance. You can read off the set speed by compa-
ring the numbers on the adjusting wheel (2) and the speed table
(see chapter: Speed Table).
By using the adjustment wheel (2) the speed may be infinitely
varied to meet the requirements of the medium to be processed.
The no-load speed of the T 10 basic driving unit may be set
between 8.000 and 30.000 1/min.
The unit is ready for service when the
mains plug has been plugged in.
The driving unit is switched on by means of the rocker switch (1).
012009

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Werke t 10 basic35 197 00

Table des Matières