Guidance and Manufacturer's Declaration—Electromagnetic Immunity
The FreeStyle Comfort is intended for use in the electromagnetic environment specified below. The customer or the user of the
FreeStyle Comfort should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electromagnetic
environment –
±8 kV contact
guidance
±15 kV air discharge
IEC 61000-4-2
±2 kV for power
supply lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/
output lines
±1 kV line(s) to line(s)
±2kV common mode on
Surge
AC lines
IEC 61000-4-5
±1kV differential on
AC lines
±2kV common mode on
outdoor I/O lines
0% U
45°, 90°, 135°, 180°, 225°,
270°, 315°)
Voltage dips, short
0% U
interruptions and
voltage variations on
power supply input
70% U
lines IEC 61000-4-11
for 25/30 cycles (0
0% U
cycles (0
Power frequency
(50/60 Hz) magnetic
3 A / m
field IEC 61000-4-8
NOTE U
is the A.C. mains voltage prior to application of the test level.
T
NOTE In some cases, the Freestyle Comfort may reset after a significant Electrostatic Discharge, and may require a Full Power
Cycle to resume normal operation. See Troubleshooting section for more details.
Compliance level
±8 kV contact
±15 kV air discharge
±2 kV for power
supply lines
N/A
±1 kV line(s) to line(s)
for 0.5 cycle (0°,
0% U
T
45°, 90°, 135°, 180°,
225°, 270°, 315°)
0% U
for 1 cycle (0
0
)
T
(30% dip in U
)
70% U
T
T
)
for 25/30 cycles (0
0
for 250/300
0% U
T
)
cycles (0
0
3 A / m
FreeStyle Comfort
Electromagnetic environment –
guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a
typical professional healthcare facility and
home healthcare environments.
Mains power quality should be that of a
typical professional healthcare facility and
home healthcare environments.
for 0.5 cycle (0°,
T
Mains power quality should be that of a
typical professional healthcare facility and
home healthcare environments. If the user
for 1 cycle (0
0
)
T
of the FreeStyle Comfort requires continued
operation during power mains interruptions,
(30% dip in U
)
it is recommended that the FreeStyle
T
T
)
0
Comfort is powered from an uninterruptible
power supply or a battery.
for 250/300
T
)
0
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a professional healthcare facility and
home healthcare environments.