CAUTION: When using the Oxygen Concentrator in an
automobile, boat, or on other DC sources with the DC
power supply, make sure that the vehicle is started
and running before connecting the Oxygen Concen-
trator. Failure to follow these instructions can result in
the power supply not supplying power to the Oxygen
Concentrator.
CAUTION: When the automobile in which you are using
the Oxygen Concentrator unit is turned off, disconnect
and remove the device from the automobile. Do not
store the Oxygen Concentrator in a very hot or cold
automobile or in other similar, high- or low-tempera-
ture environments.
Note: Portable and mobile RF communications equipment
can effect medical electrical equipment.
CAUTION: DO NOT leave the Oxygen
Concentrator or the Power Supply plugged
into the vehicle if the ignition is in the OFF
position. Doing so may drain the vehicle's
battery.
CAUTION: Operating or storing the Oxygen Concen-
trator outside of its normal operating temperature
range and ambient pressure range (altitude) can affect
performance and decrease battery run time and/or
increase battery charge time. Refer to the Specifica-
tions section in this manual for storage and operating
temperature limits.
CAUTION: For Oxygen Concentrators equipped with
batteries: Store in a cool and dry location to help
ensure the longevity of your battery. Storing your Oxy-
gen Concentrator for extended periods of time at high
temperatures or with a fully charged or completely
discharged battery can degrade its overall battery
life. Do not attempt to open the battery; there are no
serviceable parts inside the battery. Keep batteries
away from children.
CAUTION: ONLY USE the Manufacturer's provided
batteries. For proper battery disposal, contact your
Equipment Provider or your local government agency
for disposal requirements.
FreeStyle Comfort
Note: Ensure the cannula is fully inserted and secure. This
ensures that the Oxygen Concentrator can properly detect
inspiration for oxygen delivery. During inhalation, you
should hear or feel oxygen flow to the prongs of the nasal
cannula.
Note: Lithium batteries may permanently lose capacity
when exposed to extremely hot temperatures with the bat-
teries fully charged or completely depleted. For extended
storage, it is recommended that batteries be charged 25
to 50% and remain within a temperature range of -4°F –
68°F (-20°C – 20°C).
Note: The use of some oxygen administration accessories
not specified for use with this oxygen concentrator may
impair its performance. Recommended accessories are
referenced within this manual.
Note: Do not operate the Oxygen Concentrator without the
air intake filter in place. If a second filter is provided, insert
the replacement filter before you clean the dirty filter.
Clean the dirty filter in a warm soap and water solution,
then dry thoroughly prior to use.
Note: The Manufacturer does not recommend the steriliza-
tion of this equipment.
Note: There is never a danger of depleting the oxygen in a
room when you use your oxygen concentrator unit.
WARNING: THE OXYGEN CONCENTRATOR
SHOULD NOT BE OPERATED OR PLUGGED
IN TO EXTERNAL (AC/DC) POWER WITHOUT
A BATTERY INSTALLED. DO NOT ATTEMPT
TO POWER ON THE UNIT OR CONNECT IT
TO EXTERNAL POWER WHEN THE BATTERY
PACK IS REMOVED.
ONLY REMOVE THE BATTERY PACK WHEN THE UNIT IS
POWERED OFF AND UNPLUGGED FROM THE POWER
SUPPLY. DO NOT ATTEMPT TO REMOVE THE BATTERY
PACK WHILE THE UNIT IS STILL ON OR PLUGGED INTO
THE POWER SUPPLY.
Note: If the oxygen concentrator has been stored for an ex-
tended period of time outside of its normal environmental
operating conditions shown in the table above, it should be
allowed to return to normal operating temperature before
being powered on.