Table 6: Transducer Surface Temperatures IEC 60601-2-37 (Internal Use)
Standard
42.3(a) 1, Test Method B
(IEC 60601-2-37,
Amendment 1)
42.3(a) 2
(IEC 60601-2-37,
Amendment 1)
a.
The ambient temperature shall be 23
b.
Ibid.
C8
ICT
Measured
Measured
tempera-
tempera-
ture T° C
ture T° C
2.4° C rise
3.7° C rise
10.8° C rise
11.9° C rise
°
°
C ± 3
C
Temperature Limit
Allowed max. temperature T° C
6° C from initial TMM
measurement.
Temperature rise from ambient
temperature
a
of the test object
(tissue mimicking material or
TMM) to maximum temperature
on the test object within 1 mm
of the active surface of the
transducer.
27° C rise from ambient
Temperature rise from ambient
b
temperature
, as measured on
the active surface of the
transducer, over a 30 minute
time period.