The system is intended for use in the electromagnetic environment specified below.
Table 2: Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Electrostatic
2.0KV, 4.0KV, 6.0KV
Discharge (ESD)
contact
IEC 61000-4-2
2.0KV, 4.0KV, 8.0KV air
Electrical fast
2KV on the mains
Transient burst
1KV on signal lines
IEC 61000-4-4
Surge
0.5KV, 1.0KV, 2.0KV on
IEC 61000-4-5
AC power lines to
ground
0.5KV, 1.0KV on AC
power lines to lines
Voltage dips,
>5% U
short
(>95% dip in U
interruptions
0.5 cycle
and voltage
40% U
variations on
(60% dip In U
power supply
cycles
input lines
70% U
IEC 61000-4-11
(30% dip in U
cycles
>5% U
(>95% dip In U
Power
3 A/m
Frequency
Magnetic Field
IEC 61000-4-8
4
Chapter 2: Safety
Compliance Level
2.0KV, 4.0KV, 6.0KV
contact
2.0KV, 4.0KV, 8.0KV
air
2KV on the mains
1KV on signal lines
0.5KV, 1.0KV, 2.0KV
on AC power lines
to ground
0.5KV, 1.0KV on AC
power lines to
lines
>5% U
T
) for
(>95% dip in U
T
for 0.5 cycle
40% U
T
) for 5
(60% dip In U
T
5 cycles
70% U
T
) for 25
(30% dip in U
T
25 cycles
>5% U
T
) for 5s
(>95% dip In U
T
for 5s
3 A/m
Electromagnetic Environment
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the SonoSite
ultrasound system requires continued
T
operation during power mains
) for
T
interruptions, it is recommended that
the SonoSite ultrasound system be
T
powered from an uninterruptible power
) for
supply or a battery.
T
T
)
T
If image distortion occurs, it may be
necessary to position the SonoSite
ultrasound system further from sources
of power frequency magnetic fields or
to install magnetic shielding. The power
frequency magnetic field should be
measured in the Intended installation
location to assure that it is sufficiently
low.