Maintenance
If grinding tools wear out over the course of time, espe-
cially as a result of constant grinding of very hard materi-
als, so that rapid grinding with the appropriate quality is
no longer possible. The worn out parts must be replaced.
For replacing grinding tools, see section "Exchanging
Cleaning
Disconnect the mains plug prior to
cleaning!
Dirty parts in the pulverization area can be cleaned with a
cleaning brush or dry cloth.
If there is heavy accumulation of dirt, the grinding tool and
grinding pot can be disassembled and cleaned in a dish-
washer.
Disassemble the grinding tool and container as following:
- Disassemble the grinding tool (see "Exchanging and
- Disassemble the grinding pot with socket wrench SW 17.
Then, grinding pot and grinding tool can be disassembled
and cleaned with a dishwasher or by hand.

Maintenance and cleaning

".
" section).
SW 17
Fig. 11
Note: The chamber cap cannot be cleaned in a dishwasher.
After cleaning, all parts can be assembled in reverse se-
quence.
Note: Please pay attention to O-rings during all process.
When O-rings wears out, replace them timely.
For dust-free and avoid cross contamination, we recom-
mend to replace the felt ring in the bushing nut after every
grinding with help of a small screw driver.
Hex. nut
Grinding tool
Bushing nut
Grinding pot
O-ring
O-ring
Fig. 12
Fig. 13