5. Commissioning, operation

5.1 Mechanical mounting

CAUTION!
Improper mounting
Transducer is damaged.
Remove the protection cap not until shortly before mounting.
Protect the case from any forces being applied (e.g. lever force).
Do not cross the threads.
Tools required
Flats 16 and flats 19 open-ended spanner
Screwdriver 1 ... 1.5 mm
Scissors
Set of hexagon wrench keys (for WUD-26)
Model WUD-20, single end
Tighten the transducer using the spanner flats.
The correct torque depends on the dimensions of the process connection and the gasket used
(form/material). Observe the manufacturer's instructions for the connections.
Model WUD-25, flow through
1. Tighten the union nut hand-tight and align the instrument.
2. Tighten the union nut.
Welding Connections (only WUD-20, WUD-25)
The weld needs to be fully penetrating, but amperage and heat need to be minimised. We recommend
flowing Argon gas through the transducer during welding. This will help to cool the transducer. Prior to
welding tubing to the transducer, it is recommended that a few test welds be made.
WIKA operating instructions Ultra High Purity transducer, model WUD-2x
EN
13