HEINE Optotechnik only approves the agents and procedures mentioned below.
Other procedures/reprocessing agents have to be validated.
Cleaning and disinfection may only be carried out by personnel with sufficient
hygienic knowledge.
Observe the instructions of the disinfectant manufacturer.
Do not use spray or immersion disinfection, dripping wet or heavily foaming
tissues. Do not use automated or ultrasonic reprocessing.
Procedure
Clean and disinfect the retinometer manually (wipe cleaning and wipe disinfection).
Recommended agents:
Cleaning agent: Neodisher
Disinfectant agent: alcoholic (e.g. Incides
Changing the light source
Verify that the lamp voltage complies with the supply voltage of the handle.
Remove the LAMBDA 100 from the HEINE handle. Unscrew the connecting piece
and carefully remove the bulb (5).
Wipe the dome of the new bulb with a clean cloth and place the bulb in the
retinometer.
Screw the connecting piece back on.
Maintenance and Service
The instruments do not require maintenance or service.
General Warnings
Check the correct operation of the device before use! Do not use the device if
there are visible signs of damage or the light begins to flash.
Do not use the device in fire- or explosive risk area (e.g. oxygen saturated or
anaesthetic environments).
Do not modify the device.
Use only original HEINE parts, spare parts, accessories and power sources.
Repairs shall only be carried out by qualified persons.
General Notes
The warranty for the entire product is invalidated if non-genuine HEINE products
or non-original parts are used and if repairs or modifications are made to the
device by persons not authorized by HEINE. For more information, please visit
www.heine.com.
If you don't use the device for a longer period of time, please remove the batteries
in advance.
Disposal
The product must be recycled as separated electrical and electronic devices.
Please observe the relevant state-specific disposal regulations.