Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an
environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Transistor/Electrical
fast transient
IEC 61000-4-4
Impulses
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
IACER Srl
EN 60601-1-2
Compliance
Test level
± 8kV
± 8kV contact
contact
± 15kV air
± 15kV air
± 2kV for
± 2kV per
power
power
supply lines
supply lines
±
1kV
for
±
differential
differential
mode
mode
±
2Kv
for
±
common
common
mode
mode
0% U
0% U
T
per
per 0.5 cycles,
cycles,
differential
differential
angles
angles
0% U
0% U
T
per 1 cycle
per 1 cycle
70% U
70% U
T
per
25/30
per
cycles
cycles
0% U
0% U
T
per 5s
per 5s
3A/m
28
Electromagnetic
level
environment –guidance
Floors should be wood,
concrete or ceramic made.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
1kV
for
Mains power quality should
be that of a typical
commercial or hospital
2Kv
for
environment.
T
Mains power quality should
0.5
be that of a typical
commercial or hospital
environment. If the user of
the device requires
T
continued operation during
power mains interruptions,
T
it is recommended that the
25/30
device be powered from an
uninterruptible power
T
supply or a battery.
Power frequency magnetic
fields should be at levels
3A/m
characteristic of a typical
location in a typical
commercial or hospital
MNPG326