TROUBLE-SHOOTING
Switch off the device before any procedure and unplug the power cable from the socket.
PROBLEM
The device does not work
Lack of suction
Lack of suction caused by fluid
leakage
Blocked float
Vacuum power poor and/or
non-existent
If, after checking the aforementioned conditions, the device still does not work properly, we recom-
mend that you contact your local dealer or an authorised FLAEM service centre.
CAUSE
- Battery flat
- Battery not inserted properly
- The power cable has not been
inserted properly into the device
socket or into the mains power
socket
The collection cannister lid has
not been correctly positioned on
the cannister
Lid gasket not in place
Blocked filter
Incrustation on the float
Vacuum adjustment knob com-
pletely open
Blocked protection filter
Connecting tubing to the filter
and device clogged, bent or
disconnected
Cannister lid overflow valve
closed or blocked
Dirty, blocked or damaged
pump
19
SOLUTION
- Charge the battery
- Check that the battery connec-
tor is inserted properly
- Correctly insert the power ca-
ble into the sockets
- Correctly position the collec-
tion cannister lid
- Correctly position the gasket
on the lid
- Replace the filter
- Remove the cannister lid and
safety device and extract the
float. Continue by carrying out
the cleaning procedures as
described in the paragraph,
"CLEANING, SANITISATION,
DISINFECTION, STERILISA-
TION"
- Fully close the adjustment
knob and check the vacuum
power
- Replace the filter
- Check the condition of the
tubing, replace it if blocked and
correctly connect it as per the
"ASSEMBLY DIAGRAM" of this
manual
- Unblock the overflow valve,
keeping the device upright
- Take the device to your trusted
dealer or an authorised FLAEM
service centre