Led Functions; Replacing The Battery - FLAEM Aspira go Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22
ATTENTION: periodically check the battery status so that you are never in a situation requiring the device
urgently without being able to rely on the internal power source
2.
Operation and charging with the vehicle power outlet (12V DC), or with the multi-voltage power
supply:
2.1. Vehicle power outlet cable (12V DC) (20):
2.1.1. Using the vehicle power outlet cable (20), connect the external device socket (18) to the vehicle
power outlet. Check the battery charge status of the vehicle or boat before connecting the power outlet
cable.
2.1.2. Switch the device on by acting on the switch (2); the operation is displayed by the activation of the
green LED (1a). The device is designed for an intermittent use of 30 min ON/30 min OFF.
2.2. Multi-voltage power supply (switching) (19).
2.2.1. Connect the connector to the external device socket (18). Plug the power cable into the mains
socket that is compatible with the device voltage. The position of the socket must be such that the
device can be easily unplugged from the mains network. In the event that the power cable plug is unsuit-
able for the mains socket, contact the dealer or an authorised service centre.
2.2.2. Repeat the instructions in section 2.1.2
2.3. Charging the internal battery using the vehicle power outlet cable or multi-voltage power supply:
Check the battery charge status of the vehicle or boat before connecting the power outlet cable.
2.3.1. The battery is charged by leaving the vehicle power outlet cable or the multi-voltage power supply
(19) connected to the external device socket (18). The activation of the green LED (1a) and of the yellow
LED (1b) indicate that the internal battery is charging.
Charge the device for 24 hours the first time it is used; subsequently 7 hours (minimum) are sufficient to
charge model P1611EM-20 and 4 hours (minimum) for model P1611EM-30.

LED FUNCTIONS

1a 1b 1c

REPLACING THE BATTERY

To replace the battery, follow the instructions on page 1. Release, lift and remove the lid of the battery com-
partment located at the back of the device (dwg.1,2,3); remove the battery to be replaced and insert the new
one, fitting the connector properly (dwg.4,4a); put the battery compartment lid back in place and close it
(dwg.5,6). Then charge the battery following the instructions in section 2.3.1.
Flat batteries must be disposed of in special waste collection containers, or by contacting a suitable waste
disposal centre.
It is forbidden to use batteries other than those provided with the medical device. If necessary contact an
authorised FLAEM service centre.
GREEN LED (1a): when it is switched on it means that the external power
supply is connected (12VDC), i.e. the switching power supply has been
inserted in the device socket (18).
YELLOW LED (1b): when it is switched on it means that the battery is charg-
ing while when it is switched off it means that the battery is charged and
ready for use*.
Only for model P1611EM-30: of the YELLOW LED (1b) flashes is means that
the battery is not connected and/or is not connected to the circuit properly
(see dwg. 4, 4a, 4b).
RED LED (1c): flashing and acoustic alarm active, both mean that the battery
is about to run out and must be charged as soon as possible.
*IMPORTANT: to charge the battery insert the power supply plug (12VDC)
into the device socket (18).