3. Specify the data selection parameters. Push the left and right arrow keys to make a selection.
Push the up and down arrow keys to select an option.
Option
Result type
Application
Date
Operator
4. If View data log was the selected option, push View to see the selected data.
• Push the up and down arrow keys to select a line of data and push Detail to see more data
• If only one Application is selected push the up and down arrow keys to select a line of data
and push Detail to see more data, or push the left and right arrow keys to see related graphs
5. If Export data log or Delete data log was the selected option, push Preview to see the selected
data, then push Export or Delete to start the procedure.
8.5 Purge
Use this procedure to remove air bubbles from the system. Refer to
measurement
on page 20 for instructions.
8.6 Cell management
Use this option to fill or empty the measurement cell with the pumps. These procedures stop
automatically after a pre-determined time. The time remaining is shown against Security time on the
display.
1. From the home screen, select Cell management.
2. Push the left and right arrow keys to make a selection.
Option
Fill cell (pump
1)
Empty cell
(pump 2)
Stop pumps
Stir
3. Push Exit to go back to the home screen.
Section 9 Maintenance
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
26 English
Description
Sets the type of result available.
Sets the available applications.
Sets the date range.
Sets the available operators.
Description
Use this option to fill the measurement cell with reagent from pump 1.
Chemical exposure hazard. Before selecting this option, examine the level
of liquid in the waste bottle. Make sure that there is enough space in the
bottle for all of the liquid from the measurement cell.
Use this option to empty the measurement cell with pump 2.
Use this option to stop the Fill cell and Empty cell procedures.
Use this option to start or stop the magnetic stirrer. This option can also be
used when the Fill cell and Empty cell procedures are operating.
C A U T I O N
C A U T I O N
Prepare the instrument for