TCM 3000 BL - Morcellator
Risk of injury!
The distal end of the cutting tube is extremely sharp.
Precautions:
• Handle the cutting tube particularly cautiously and carefully.
• Where possible, have the Morcellator with protective tube in protected position (NO CUT), particularly
if the cutting tube is not used for any period.
• Insert and remove the protective tube and cutting tube under visual control (endoscopy).
Risk of injury!
When inserting the Obturator through the percutaneous access, there is risk of injury to blood vessels,
intestinal loops or the bladder. This can result in significant complications, which will require a Laparoto-
my.
Risk of damage to the blade of the cutting tube!
If the jaws of the gripper are not completely closed when it is pulled from the protective tube, the blade
will be damaged (see pictures).
Risk of injury!
Damaged blades and cutting tubes must not be used.
Cutting tubes must not be sharpened.
22
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16