Nouvag TCM 3000bl Mode D'emploi page 58

Masquer les pouces Voir aussi pour TCM 3000bl:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 85
9.
Connecting pedal
Align the red dot on the device connection plug of the
pedal with the red dot on the pedal connection of the
control unit.
Plug the pedal device connection plug into the pedal
connection of the control unit.
The plug must lock in position.
Functional check
Check Morcellator for loose parts
Check that Morcellator is correctly assembled
Are all labels easily readable?
Check cutting tube: cutting tube evenly ground, no broken edges or bends?
Switch on control unit
Switch on motor (press pedal).
Run motor with Morcellator for about 20 seconds; it must reach the set speed.
If it does not run evenly, switch off motor with Morcellator.
Switch off control unit.
The control unit will only operate if the pedal has been connected to the control unit.
Check that Morcellator is correctly assembled.
Check that the gear unit runs smoothly; if necessary treat it with Nou-Clean care spray.
Check that the motor is clean and runs smoothly; if necessary clean it and treat with Nou-Clean care spray.
Repeat the function check.
Risk of injury by motor failure
Problem: the motor becomes very hot. Cause: the motor was damaged during preparation.
Remedy: do not continue to use the motor. Always have a spare motor available.
14
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
TCM 3000 BL - Morcellator

Publicité

Table des Matières
loading

Table des Matières