TCM 3000 BL - Morcellator
3.4

During use

The device is not sterile on delivery. All
sterilizable parts must be sterilized before
use (see Chapter 7 Instrument prepara-
tion).
Operate the device only outside the dan-
ger zone of explosive, flammable mixtures
or gases.
The motor ventilation slots must be kept
clear in order to prevent the motor from
overheating.
Ensure that the operating voltage setting
corresponds to the local mains voltage.
The risks and benefits of using tissue
extraction bags to capture tissue released
during laparoscopic power morcellation
procedures remain to be determined.
Instrucions for use for TCM 3000 BL - Morcellator, Ref. 31946, V26/16
Never operate the clamping mechanisms of the cut-
ting tubes while the system is running. This could
result in instrument damage.
Keep hands clear of the cutting tubes when operat-
ing the device. Danger of serious injury.
The TCM 3000 BL - Morcellator must be operated
under the continuous supervision of medically quali-
fied persons only.
Disregarding the guidelines on intermittent opera-
tion (1 min on, 1 min off for 8 cycles) may cause burns
at contact with the gear unit.
EN
7