Ending An Incubation From The "Main" Menu; Switching The Heating Off; Switching The Shaker Off (Speed 0 Rpm); Switching The Working Illuminance, Photo Illuminance And Cooling On Or Off - Sartorius Stedim Biotech CERTOMAT CTplus Mode D'emploi Original

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5.3
Editing individual values of an incubation from the
"Main" menu
While the incubation is running, changes can be made
to individual values (temperature, speed) or states
(working illuminance, photo illuminance, cooling) at
any time.
A running incubation process does not have to be
stopped beforehand in order to do so.
A speed entry of "0" rpm does not completely stop the
shaker drive!
The drive continues to run at approx. 5% of the max.
speed.
The changing of values and states is carried out
analogously to the entries described in chap. 5.2.1 to
5.2.7.
5.4

Ending an incubation from the "Main" menu

In order to end an incubation from the "Main" menu,
proceed as follows.

5.4.1 Switching the heating off

- Press the "TEMP" key.
- The input menu of the temperature regulator
appears.
- Set the "Mode" to "off" and confirm with "ok".

5.4.2 Switching the shaker off (speed 0 rpm)

- Press the "SHAKER" key.
- The input menu of the speed controller appears.
- Set the "Mode" to "off" and confirm with "ok".
5.4.3 Switching the working illuminance, photo
illuminance and cooling on or off
The switch states for working illuminance, photo
illuminance and cooling are changed in the same way
as sections 5.2.3 to 5.2.7.
26
The heating is switched off.
The "TEMP" key shows the grey
underline for operating mode
"off". The entered temperature
(37 °C) remains stored.
The shaker drive is switched off.
The "SHAKER" key shows the
grey underline for operating
mode "off".
6

"Controller" function

After pressing one of the main function keys
(TEMP, SHAKER, CO
respective controller appears.
The respectively selected function (TEMP, SHAKER, CO
HUMID) is accessed with the "Controller" key.
"Controller" key
Fig. 25: Input menu for "TEMP controller"
Main functions for accessing the "Controller" menu
The following main functions provide access to the
"Controller" menu:
Key, Symbol
Meaning, Use
- Set point controller for
- Set point controller for shaker
- Set point controller for CO
- Set point controller for humidity
, HUMID) the input menu of the
2
temperature in the CERTOMAT®
CTplus
- Works with the temperature of
the CERTOMAT® CTplus as a
control parameter, measured
with the sensor that is
mounted in the air stream
speed in the CERTOMAT® CTplus
- Controls the shaker speed of
the CERTOMAT® CTplus
concentration in the CERTOMAT®
CTplus
- Controls the CO
concentration
2
in the incubation cabinet of
the CERTOMAT® CTplus
set point in the CERTOMAT®
CTplus
- Controls the humidity set point
in the incubation cabinet of
the CERTOMAT® CTplus
,
2
2

Publicité

Table des Matières
loading

Table des Matières