A leak test and pressure testing shall be performed after assembly has
been successfully completed (see the sections entitled „Leakage test").

Leakage test

Perform a leakage test before every combustion process.
Always check the lid is properly sealed with the decom-
position vessel after carrying out maintenance.
Follow the safety instructions.
• Fill the decomposition chamber in the calorimeter system or using the
C 248 oxygen filling station.
• Remove the decomposition vessel and immerse fully in a container
filled with water.
• Check the decomposition vessel for oxygen leaks by watching for bub-
bles.
• If bubbles are seen then the decomposition vessel is not sealed cor-
rectly.
• Vent the decomposition vessel and check the seals.
Never use a leaking decomposition vessel in the calorimeter!
Degase
For the C 5000, venting of the decomposition vessel takes place in the
device itself; for the C 2000 and C 200 the venting button supplied with
the device is required.
The vent knob is supplied for
manual release of tension in the
decomposition vessel. (Even in
case ot error function of the filling
unit in the calorimeter C 2000 / C
5000).
Escaping
decomposition gases
might be detrimental
to health.
Decomposition vessel identification
The individual parts of the decomposition vessel are clearly identified by
means of the engraved serial number and the technical data.
The individual parts of several decomposition vessels (cover, union nut
and bottom part) are not interchangeable.
lfd. Gerätenr.
serial no.
(Chargen-Nr.)
Bottom
Jahr
year
!
Cover
Betriebsdruck
operating pressure
(Chargen-Nr.)
Betriebstemperatur
operating temperature
lfd. Gerätenr.
Prüfdruck; Prüfdatum
serial no.
Innenvolumen
test pressure; Test date
inner volume
26
If one of the parts indicated becomes damaged then the
complete decomposition vessel has to be sent to the factory
for testing.

C 5010 Technical data

Operating pressure
max. 230 bar (23 MPa)
Test pressure
330 bar (33 MPa)
Volume
260 ml
Nuclear temperatures
max. 1300 °C
Sample decomposition
isochor
Oxygen valve
plate valve
(vacuum operation not possible)
Dish holder
as in DIN 51 900
Material in contact with the product
1.4439
2.4610.40

C 5010 Spare parts list

Pos. 11 Spring ring
Pos. 25 Crucible holder
Pos. 12 Compression ring
Pos. 27 Deflector
Pos. 13 Sealing ring
Pos. 28 Shaft C 5010
Pos. 15 Locking screw
Pos. 30 Hexagonal nut M4 A2
Pos. 16 Valve piston
Pos. 31 Hexagonal nut M3 Hastelloy
Pos. 17 O-ring 3.1 x 1.6
Pos. 40 Lable set
Pos. 20 Bushing
Pos. 41 Ignition wire
C5010.3
Pos. 21 O-ring 4 x 1.5
Pos. 1061 C4 quarz crucible
Pos. 22 Insolating bush
Pos. 45 Grip
Pos. 23 Female insert contact
Pos. 49 emergency-vent knob
Pos. 24 Ignition electrode

C 5010 Spare parts diagram

Align with
an Nut
11
ausrichten
groove
23
12
22
13
21
17
21
16
20 24 41 43
1061
15
30
27
31
28
40
45
31
25
49
27

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C 5012

Table des Matières