Ce Declaration Of Conformity; Explanation Of Symbols - IKA C 5010 Mode D'emploi

Table des Matières

Publicité

Source language: German
EN
Contents
Page

CE Declaration of conformity

Warranty

Explanation of symbols

Safety instructions
C 5012 Spare parts list
CE Declaration of conformity
We declare under our sole responsibility that this product corrosponds to
the regulations for pressure vessels 2014/68/EU.
A copy of the complete EU Declaration of Conformity can be requested
at sales@ika.com.
16
16
16
Warranty
17
19
20
In accordance with IKA warranty conditions, the warranty period is 24
20
months. For claims under the warranty please contact your local dea-
21
ler. You may also send the machine direct to our factory, enclosing the
21
delivery invoice and giving reasons for the claim. You will be liable for
22
freight costs.
22
The warranty does not cover worn out parts, nor does it apply to faults
22
resulting from improper use, insufficient care or maintenance not carried
22
out in accordance with the instructions in this operating manual.
24
24
24
Explanation of symbols
24
25
25
This symbol identifies information that is absolutely essen-
26
tial for your safety and health. Failure to observe this in-
26
formation may result in negative effects for your health and
26
injuries.
27
27
27
28
28
28
Safety instuctions
Read the operating instructions in full before star-
ting up and follow the safety instructions.
• Keep the operating instructions in a place where they can be accessed
by everyone.
• Ensure that only trained staff work with the appliance.
• Follow the safety instructions, guidelines, occupational health and sa-
fety and accident prevention regulations.
Intended purpose:
The C 5010 and C 5012 decomposition vessels must only be used for de-
termining the gross calorific value of solid and liquid combustible mate-
rials in accordance with DIN 51900, BS 1016 T5, ISO 1928, ASTM 5468,
ASTM 5865, and ASTM 4809.
Operation of decomposition vessels is only permitted in combination
with the IKA calorimeter C 200, C 2000 or C 5000.
Operating requirements:
The maximum amount of energy input into the decomposition vessel
must not exceed 40000 J. (Select the weight of the sample accordingly).
The permissible operating pressure of 230 bar (23 MPa) must not
be exceeded. The maximum permissible operating temperature must
not exceed 50 ° C.
Do not fill the decomposition vessel too full of the sample. Only fill the
decomposition vessel with oxygen up to a maximum pressure of 40 bar
(4 MPa). Monitor the adjusted pressure on the pressure reducer.
Perform a check before every combustion to ensure there
are no leaks (see the chapter entitled „Leakage test").
16
Explosive substances:
Many substances tend to combust in an explosive manner (for example
because of the formation of peroxide). This may cause the decompositi-
on vessel to burst.
The C 5010 and C 5012 standard decomposition vessels must not be
used for examinations on samples that are capable of exploding. It is
absolutely essential to use a special high-pressure decomposition ves-
sel to contain the sample in these cases!
Notes on the sample:
Substances of which the combustion behavior is not known must be
examined for their combustion behavior before combustion in the de-
composition vessel C 5010 and C5012 (danger of explosion).
If you are burning unknown samples, leave the room or keep a safe
distance between you and the calorimeter
Benzoic acid must only be burned in the form of pellets!
Combustible dust and powder must be compressed into pellets before
combustion. Oven-dry dust and powder such as wood chips, hay, straw,
etc. burn in an explosive manner! They must be moistened first! Readily
combustible liquids with a low vapor pressure must not be come in di-
rect contact with the cotton thread (for example tetramethyl dihydrogen
disiloxan)!
Combustion residue, auxiliary materials:
In addition, toxic residues of combustion are possible in the form of gas-
ses, ash or precipitates on the inner wall of the decomposition vessel,
for example
Observe the accident prevention requirements applicable to the
activity and the work station. Wear personal safety equipment.