Startup; User Interface And Navigation - Hach 9586sc Oxygen Manuel D'utilisation De Base

Masquer les pouces Voir aussi pour 9586sc Oxygen:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Install the reagents
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
1. Fill the empty sample conditioning bottle with DIPA (diisopropylamine) and enter the hazardous
reagent information on the blank label attached to the bottle.
2. Install the sample conditioning bottle on the front of the analyzer.
3. Connect the supplied tube to the measuring cell. Install the tube completely into the bottle of
conditioning solution. Make sure that the tube fittings are tight to prevent leakage of fumes from
the conditioning bottle into the atmosphere.
4. If the chemical zero cartridge is used, install the cartridge on the rear of the analyzer.

Startup

Make sure that the flow rate and pressure do not exceed the values in
1. Open the valve on the sample line to let sample flow through the analyzer.
2. Turn the knob on the flow meter to set the flow rate.
3. Examine the plumbing for leaks and stop any leaks if found.
4. Apply power to the controller.
5. Make the applicable menu selections when the controller starts.

User interface and navigation

User interface
The keypad has four menu keys and four directional keys as shown in
16 English
W A R N I N G
Specifications
on page 3.
Figure
4.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

9586sc

Table des Matières