Remove the neon one by one holding
them tightly and turning them with both
hands for two semi-snaps in one
direction POS 3. Now you can take the
neon off their base and get them out of
middle of the chamber.
Take all the 4 neon off and lean them on
a flat surface.
Clean gently the germicidal chamber's
internal walls POS 5 with a soft antistatic
duster to get them rid of dust and dirt.
Clean the UVC neon with a soft duster
before the installation.
Replace the neon when they are no
more efficient or in case the red led
lightens (FIG.05,Pos.8)
Perform the installation with the same
passages of the old neon's removal.
FIG02
Install the chamber's cover and tighten
the 2 screws on it.
In order to clean or replace the internal
filter POS 4 pull the tab and take it off
completely , then clean or replace the
cartridge filter.
Lock and take the key off the lock.
Keep a record of the time the neon have
lasted. They should be replaced at least
every 12 months. (cfr schedule 2)