Cleaning: only clean the device using a dry cloth. Do not use solvents or other chemical
substances. Remove batteries before cleaning the device.
We recommend a functional test carried out by the manufacturer every 24 months.
The manufacturer retains that the MIO-CARE device cannot be repaired by personnel external
to its own company. Any work of this nature carried out by personnel not authorised by the
manufacturer will be classed as tampering with the device, freeing the manufacturer from
responsibility for the warranty and hazards that the operator or user may be exposed to.
N.B. disconnect the cables before placing the electronic stimulator in the bag. If not, the cables
will bend excessively close to the connectors. This may damage the cables.

9. Resolution of problems and questions.

Any type of work on MIO-CARE must be carried out exclusively by the manufacturer or by an
authorised dealer. In any event, any presumed malfunction of MIO-CARE must be verified
before sending the device to the manufacturer.
Here below are some typical situations:
MIO-CARE cannot be turned on:
• Recharge the batteries and try to switch it on again; if it still does not switch on, check that
the power supply socket is working.
• Check that the ON/OFF button has been pressed correctly (held down for at least one second).
MIO-CARE does not transmit electric impulses or transmit low intensity:
• Check that the cable jacks have been inserted in the electrodes and that the plastic
protection has been removed from the electrode.
• Check that the cables have been connected correctly (connector well inserted in the device).
• Check that the electrodes have been connected correctly (according to user manual
instructions) and check that electrodes are in good condition.
MIO-CARE switches off during operation:
• Recharge the batteries and start the treatment again. If the problem persists contact the
manufacturer.
Stimulation is not comfortable:
• Check that intensity is not too high and decrease intensity if needed.
• Check the electrodes postion: follow user manual and electrodes position manual
instructions to avoid that electrodes are too close.
• Check the sized electrodes section (suggested by the manufacturer) : electrodes too small
can cause uncomfortable stimulations.
Intermittent output:
• Check that electrodes and cables have been connected correctly. If the problem persists
replace cable or contact the manufacturer.
Stimulation is ineffective:
• Check that electrodes have been positioned correctly. If the problem persists contact the
doctor/therapist.
IACER S r
48
from 30% to 85%
from 700 hPa to 1060 hPa
from -10° C to +50° C
from 10% to 90%
from 700 hPa to 1060 hPa
PG95 Rev 02 Ed 31 01 12