Dimension for actuator elements
Switch- and potentiometer actuator
Glass for measuring instrument AM/VM45
Glass for measuring instrument AM/VM72
X = fixing dimensions
1
2
Fig. 1
12
Explosion-protected control units
GHG 411 and 412
1.4
AM45 / AM72 measuring Instrument
VM45 / VM72 measuring instrument:
ATEX type examination certificate:
Marking acc. to 2014/34/EU and standard:
Moving iron
EN 60079-0
Moving coil
IECEx type examination certificate:
Category of application:
Moving iron
IEC 60079-0
Moving coil
Operating temperature
Rated voltage:
Measuring accuracy:
Overload range:
Measuring range:
AM:
VM45:
VM72:
Power consumption:
VM45:
VM72:
Coil:
Internal resistance:
Li:
Ci:
Ui:
I i :
Supply terminals:
Test torques:
Weight:
AM/VM 45
AM/VM 72
1.5

Terminal block

ATEX certificate:
Marking acc. to 2014/34/EU and standard:
EN 60079-0
IECEx certificate:
Category of application:
IEC60079-0
Type of terminal:
Rated voltage:
Conductor cross-section:
Test torques:
Weight:
1.6

Intrinsically safe circuits:

Intrinsically safe circuits: Max. safe voltage U
BVS 14 ATEX E 125
D II 2 G Ex e IIC Gb
D II 2 G Ex e mb II CGb
D II 2 G Ex ib IIC Gb
IECEx BVS 14.0082U
Ex e IIC Gb
Ex e mb IIC Gb
Ex ib IIC Gb
-55 °C
to
+80 °C
to max. 500 V
Moving iron
Klasse 2,5
sec.
10 -
fold at
- 25
sec.
25 -
fold at
- 4
sec
50 -
fold at
- 1
indicated
1:1,5
n / 1A
direct
0 - 25 A
6 - 415V
6 - 660V
0.91 - 1.76VA
0.91 - 2.65VA
-
-
-
-
2 x 0.5 - 2.5 mm²
fine- / multy wire
solid wire
1 x 4.0 mm²
2.5 Nm
approx. 0.35 Kg
approx. 0.40 Kg
SEV 13 ATEX 0178 U
II 2 G Ex eb IIC
IECEx SEV 13.0012 U
Ex eb IIC
6 x MUT 4
up to 352 V
2 x 0.2 - 4.0 mm²
0.6 - 0.8 Nm
approx. 0.08 kg
:
690 V
m
eff
Safe galvanic isolation from all other circuits
and earth
E
Moving coil
Klasse 1,5
sec
10 -
fold at
- 5
0 - 24 mA
4 - 24 mA
26.5
turns
2.5 Ohm +- 30%
max. 0.1 mH
max. 0.1 nF
max. 30 V
max. 150 mA