Cable Protection (Provided By Customer); Shortening The Cable; Cable Extension - Hbm Pw27A Notice De Montage

Table des Matières
Les langues disponibles

Les langues disponibles

Electrical connection
5.2

Cable protection (provided by customer)

Cable protection system*, comprising:
* Supplier such as Flexicon, e‐mail: info@icp‐solutions.de
Even without a cable protection system, standard load cells can achieve
IP68/IP69K degree of protection. In applications where the PUR cable of the
load cell can be chemically or mechanically attacked or destroyed, extra cable
protection can be provided by means of standard cable protection systems.
5.3

Shortening the cable

As the load cell connection uses a six‐wire configuration, you can shorten the
6‐wire load cell cable without this changing the load cell calibration.
5.4

Cable extension

Use shielded, low‐capacitance measurement cables only for extension. Ensure
that connection is perfect, with a low contact resistance.
The cable of a six‐wire load cell can be extended with a cable of the same
type.
12
Unscrew the M10 sleeve and
replace with an M10
threaded tube coupling
A05201_02_Y00_00 HBM: public
Screwed cable gland
Flexible cable
protection tubing*
PW27A
Table des Matières
loading

Table des Matières