5.2
Checking the balance with the internal weight (ABJ only)
Sequence:
(1)
Remove the items being weighed from the balance plate and press the TARE/
key ; the display will indicate zero.
-
(2)
Press the CAL/MENU key until I CAL appears.
Note: Balances with automatic internal adjustment are set at "I cal" as a
standard, i.e. set for adjustment using an internal weight. This setting can
be checked using the "Setting" menu, see chapter 4.4.4, and altered if
necessary.
(3)
Now press the TARE/
(4)
The display will subsequently indicate CAL 1 followed by CAL 0.
(5)
The display will indicate CAL END before reverting to weighing mode.
Adjustment has been completed successfully.
5.3

Checking the balance with the external weight

Sequence:
(1)
Remove the items being weighed from the balance plate and press the
TARE/
key; the display will indicate zero.
-
(2)
Press the CAL/MENU key until FUnC.SEL appears.
(3)
Then press the TARE/
TARE/
key. again ( E CAL appears ) .
-
(4)
Press the CAL/MENU key until E tESt appears, then press the TARE/
the check will start.
(5)
The zero will be flashing. Following this the weight value will flash for adjustment.
(6)
Place the necessary adjustment weight on the balance plate.
(7)
The zero weight will now be flashing on the balance.
(8)
Remove the weight from the balance plate.
(9)
If no error message is issued, the balance will display zero again, thereby
indicating that the check has been successful.
28
key
The display will indicate CAL 2.
-
.
key. The display will indicate CAL, then press
-
key
-
;