15.2 Maintenance schedule
Maintenance performed periodically or after number of samples.
Maintenance Schedule Interrliner (Example)
Sample volume: 350 per day (5 working
days)
or earlier after number of samples
Perform End-of-day wash
Clean outside needle assembly
Check tubing/diluent syringe
Clean outside instrument
Clean Fill nozzle
Clean waste separator
(p. 98)
Check sensors
Replace Fill nozzle O-ring
Replace bacterial air filter (HEPA)
Replace Disc filter (white) Only if internal
waste container is used
Replace Rinse tube assembly
Replace Saline tube assembly
Run Fill and clean
Replace waste cassette assembly
122
Daily
Weekly
Level 4
(monthly)
7.500 samples
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
Level 3
Level 2
Level 1
(quarterly)
(halfyearly)
(yearly)
23.000
46.000
91.000
samples
samples
samples
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
D0054061-FR Version C01 (17-12-2024) Interrliner
Annexe pour l'Interrliner
Parts
Total
amount/year
Per 91.000
samples
QWLV050004
12
QWLV040002
12
QWLV040001
12
ESRI090902
12
ESRI090903
12
ESRI090921
4