The SCMHV signal conditioner is used for the differen
tial measurement of high voltages with QuantumX mod
ules offering strain gauge full bridge input, for example
MX840A, MX440A and MX410/B.
Within the scope of measurement category II (supply net
work from the socket), the maximum operating voltage to
ground potential is 300 V DC or 300 V rms AC. Taking
the maximum acceptable rms value into account, the
peak value of the voltage must not at any time exceed
700 V between the connections (differential mode) and to
ground potential (common mode).
Its robustness in the face of high voltages and transients
also makes the SCMHV suitable input protection for
QuantumX modules, when the voltages to be measured
include high voltage peaks.
The 10 V measuring range allows low voltage values to
be measured at a high potential and currents to be mea
sured via an external shunt.
The SCMHV applies a safety limit to the measuring volt
age, which is electrically isolated from the adjacent chan
nels in the QuantumX module. This allows any number of
SCMHVs to be operated without mutual interference.
The signal conditioner can be automatically identified and
adjusted with TEDS (electronic data sheet).
S QuantumX SCM-HV Voltage signal conditioner
Order no.: 1-SCM-HV
S Quick Start Guide
A3615-4.0 HBM: public