P P r r e e v v e e n n t t i i v v e e m m a a i i n n t t e e n n a a n n c c e e
Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks
more often based on your level of product usage. Service only by qualified personnel.
N N o o t t e e - - Clean and disinfect the exterior of the support surface before inspection, if applicable.
O O p p e e r r a a t t i i o o n n
Cleaning and disinfecting
Inspection
S S t t a a i i r r - - T T R R E E A A D D system belt
reconditioning
Inspect the following items:
All fasteners secure (reference all assembly drawings)
All welds intact, not cracked or broken
No bent or broken tubing or sheet metal
No debris in wheels
All wheels secure and roll
Front casters are secure, roll, and swivel (if equipped)
Wheel locks secure the wheels when on and clear the wheels when off
Chair unfolds and locks
No rips or cracks in seat or backrest
Restraint straps intact and operate
Foot end carry handles extend and lock in place
Head end carry handles fold and unfold
Upper control handle extends and locks in all positions
S S t t a a i i r r - - T T R R E E A A D D system mechanism unfolds and locks in place (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt rolls (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system belt inner cords do not show; replace if necessary (if equipped)
S S t t a a i i r r - - T T R R E E A A D D system performs as desired; recondition belts if necessary (if equipped)
No lubricants present on the S S t t a a i i r r - - T T R R E E A A D D system belts or the track frame surfaces
Upper release handle cable not worn or frayed; replace if necessary
Accessories and parts intact and operate
Product serial number:
Completed by:
Date:
6254-109-005 Rev AA.0
S S c c h h e e d d u u l l e e
Each use
Inspect chair every 6 months
After usage on approximately 500
flights of stairs or if the S S t t a a i i r r - - T T R R E E A A D D
system performance is sluggish
P P r r o o c c e e d d u u r r e e
See Cleaning in the Operations
Manual
See below for checklist
See Track belt reconditioning in the
Maintenance Manual
25