4.3/ METROLOGICAL SPECIFICATIONS

All the errors are indicated as a % of Vs (value of output voltage)
- output impedance : 100 W
- zero adjustment : ± 10 A by automatic step from 25 Vo 40 mA approx.

100 A range (10 mV/A)

Intrinsic error in the field of reference
Primary current
Accuracy
Graphs of typical relative error: see appendices, page 57, 58.
Phase error (45...65 Hz)
Primary current
Phase shift
- Rise time from 10 to 90% Vs: £ 100 µs
- Fall time from 90 to 10% Vs: £ 100 µs
- Output noise:

1000 A range (1 mV/A)

Intrinsic error in the reference range
Primary current
Accuracy
Graphs of typical relative error: see appendices, page 59, 60.
Phase error (45...65 Hz)
Primary current
Phase shift
0.5...20 A
£ 1.5% + 5 mV
10...20 A
£ 3°
from DC...1 kHz
from DC...5 kHz
from 0.1 Hz...5 kHz £ 2.0 mV rms or 0.2 A rms
0.5...100 A
£ 1.5% + 1 mV
10...200 A
£ 2°
20...100 A
£ 1.5%
20...100 A
£ 2°
£ 8 mV or 0.8 A DC
£ 12 mV or 1.2 A DC
100...800 A
£ 2.5%
200...1000 A
£ 1.5°
19
100...150 A
(on DC only)
£ 2.5%
800...1000 A AC
800...1400 A DC
800..1400 A peak
£ 4%