Operating Conditions; Metrological Specifications - Chauvin Arnoux PAC 11 Notice De Fonctionnement

Pince ampèremétrique
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

4.2. OPERATING CONDITIONS

The instrument must be used in the following conditions to satisfy the safety of the user and the metrological
performance :
- Use indoors
- Working altitude :
2000 m
- Transportation altitude :
12 000 m
- Environmental conditions : see graph below

4.3. METROLOGICAL SPECIFICATIONS

All the errors are indicated as a % of Vs (value of the output voltage).
- Output impedance : 100 W
- Zero adjustment : ± 10 A by automatic step from 25 to 40 mA approx.
40 A range (10 mV/A)
„ Intrinsic error in the field of reference
Primary current
0.5 ... 40 A
40 ... 60 A
(on DC only)
Accuracy
1.5 % + 5 mV
1.5 %
13

Publicité

Table des Matières
loading

Table des Matières