• Extraordinary maintenance

Extraordinary maintenance must be entrusted to a trained
technician in possession of original spare parts:
• pay special attention to danger signals. Use goggles, glo-
ves and disposable overalls for personal protection;
• check that routine maintenance has been duly carried out
and make sure that Magnolia products are used;
• before any intervention carry out a series of washes with
Fast & Steril 3 (D), then wait 15 minutes for a complete
disinfectant action.

Every 12 months

• Check maximum detected temperatures and all alarms.
Intervene accordingly;
• check the aspirator noise level (max. 72 dB according to
the regulation 3047 (E);
• if necessary, remove dust from the circuit AC80 inside
electronic components using a blast of dry air not excee-
ding 2 bar pressure. By means of a 6 bar blast of air clean
also the small holes on the frontal cover of the aspirating
unit (15);
• check the conditions of plastic hoses, in particular the
hoses under pressure connecting the centrifugal separator
and the Hydrocyclone ISO. We suggest to replace these
hoses every 12-18 months.
Every 18-24 months
• Check the working conditions of the centrifugal separator
(11) and re-circulation valve (14).
Every 10.000 working hours
• Every time a component containing a rubber part ("O"
ring, gasket or seal) is disassembled, replace the rubber
part;
• replace the motors bearings;
• where routine maintenance is not carried out properly or
unsuitable products are used, train the staff and inform the
person in charge.